APT-5 - User’s Manual (EU) Rev. 2020-01 en
29
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
Modulation
b)
Maximum
power (W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Compliance
level (V/m)
385
380-390
TETRA 400
Pulse
modlation
b)
18
Hz
1.8
0.3
27
27
450
430-470
GMRS 460,
FRS 460
FM
c)
± 5 kHz de
viation 1 kHz
sine
2
0.3
28
28
710
745
780
704-787
LTE Band 13,
17
Pulse mod
lation
b)
217 Hz
0.2
0.3
9
9
810
870
930
800-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
28
1720
1845
1970
1 700 -
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
28
2450
2 400 -
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
28
5240
5500
5785
5 100 -
5 800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
9
NOTE:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m
test distance is permitted by IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.