APT-5 - User’s Manual (EU) Rev. 2020-01 en
26
Declaration - electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group1
Class B
Electromagnetic environment guidance The
APT-5
uses RF energy only for its internal funtion.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Harmonic emissions
IEC 61000-3-2
Class A
The
APT-5
is suitable for use in all establishme including
domestic establishments and those directly connected to
the public low voltage power supply network that supplies
buildings used for domestic purposes.
Voltage Fluctuations
And Flicker IEC 61000-
3-3:2013
Complies
Declaration - electromagnetic immunity
IMMUNITY test IEC 60601 test
level
Compliance level
Electromagnetic envronment
– guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
2, 4, 8, 15kV air
8 kV contact
2, 4, 8, 15kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst IEC
61000-4-4
2 kV for power
supply lines
1 kV for input/
output lines
2 kV for power
supply lines
N/A
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
2 kV Signal input/
output) to earth
1 kV line(s) to line(s)
2 kV line(s) to earth
N/A
Mains power quality should be that
of a typi-cal commercial or hospital
environment.
Voltage dips, short
interru tions and
voltage variations
on power supply
input lines IEC
61000-4-11
0% UT; 0.5cycle
at 0°, 45°, 90°,
135°,180°, 225°,
270° and 315° 0%
UT; 1cycle and 70%
UT; 25/30 cycles
Single phase at 0°
0% UT; 250/300
cycle
0% UT; 1cycle and
70% UT; 25/30 cycles
Single phase at 0° 0%
UT; 250/300 cycle
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the APT-5 requires continued
opera-tion during power mains
interruptions, it is rec-ommended
that the APT-5 be powered from
an uninterruptible power supply or
a battery.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
30 (A/m)
30 (A/m)
Power frequency ma netic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
NOTE:
UT is the a.c. mains voltage prior to application of the test level.