2.5.4 Membrane Switch Panel and Display
Switch ON the controller and check the
Liquid Crystal Display
(LCD) for
sharpness, legibility and brightness. If the display appears dim or segments
are missing, dismantle the display assembly and clean the contacts to the LCD
(see Section 4.13). If after reassembling the controller the display is still
defective, then replace the
zebra elastomer
and the
LCD bezel (snap fit)
only.
Use the following tests to check that the buttons on the
membrane switch
panel
are working effectively. If a button has to be pressed hard or more than
once in order to get a response, then replace the switch panel (see Section 4.5).
Press the RUN . STOP Patient LEFT and the RUN . STOP Patient RIGHT
buttons and check that the air hose and foot icons appear. Ensure the
controller delivers an impulse to each channel. Stop the impulsing by
pressing the RUN . STOP buttons.
Use the PRESETS 1-2-3 button to select each setup in turn, and check that the
information displayed matches the description in the
Operator's Instruction
Manual
.
Use the "+" and "-" buttons to alter the impulse pressure.
Press the CYCLE button, then use the "+" and "- " buttons to alter the cycle time.
2.5.5 Air Hose Sockets
Replace the
‘O’ ring seals
within the air hose sockets (see Section 4.2).
Examine the
socket face
for damage such as chips and cracks. If there is any
visible damage replace the socket (see Section 4.3).
2.5.6 Compressor
Service Interval
Compressors should be tested and maintained at 3000-hour intervals, except
for compressors identifiable by a green dot on the air filter plate, which have a
6000-hour interval.
A compressor service kit
is available containing all necessary parts to upgrade
to a 6000-hour service interval.
Compressor Performance
PC test method
The compressor output pressure may be tested by using the
Service Solutions
CD –ROM
FrontLite® test program and a PC connected through the
controller communication port using the supplied cable and
Standard test
load (1 litre)
, following the supplied instructions.
Manual test method
To manually test the compressor output pressure open the controller case
front approximately one inch and carefully remove the 7 way flat flexible
cable from the
processor PCB assembly
. Position the controller onto its
handle, on a flat surface. Disconnect the compressor supply tube at the