Insert the new power inlet in the case, with the central ground terminal to the
bottom of the case, making sure that it clips into place. The correct orientation is
essential to allow the power cord to be inserted so that it can be secured by the power
cord retainer.
Push the three connectors back on with the polarity as indicated by the markings on
the rear of the socket (see Figure 10).
Insert the power cord into the socket, and secure with the power cord retainer, screw
and washer.
Close the case.
Perform the controller safety tests to confirm the ground resistance and leakage
current (see Sections 2.2.11 and 2.2.12).
Perform the complete controller functionality tests (see Section 2.2 or 2.3).
4.18 REPLACING A FUSEHOLDER
Remove the
fuse
from the holder.
Lift up the
insulation cover
around the contacts of the
fuseholder
.
Cut off the heatshrink tubing, unsolder the two connecting wires and pull them
clear. Unscrew the plastic retaining nut on the stem of the fuseholder, and remove
the fuseholder from the outside of the case.
Insert the new fuseholder and secure it with the plastic nut.
Fuseholders
Ground test screw
Figure 11: View of fuseholders and ground test screw..