11
Key to Recommended Service Schedule
M
Mandatory Replacement
U1
Upgrade Possible: Replacement of Front cylinder housing AV6532-00 with AV6532-01 to
extend compressor service interval to 6000-hours
U2
Upgrade Possible: Replacement of Rear cylinder housing AV6708-00 with AV6708-01 to
extend compressor service interval to 6000-hours
U3
Upgrade Advised: Replacement of Compressor lubricant AV6709-00 with AV6709-01 for
improved lubrication.
U4
Upgrade Possible: Replacement Reservoir. Please call the Kendall Service Center for details.
U5 Upgrade Possible: Replacement EPROM available for upgrade to the latest software version.
U6
Upgrade Recommended: Replacement Fan AV6517-00, or –01 with AV6517-02 for
improved durability.
D
Replace on discretionary basis
L
Lubricate with ‘O’ ring lubricant AV6545-00
A
Replace using Loctite 401 after cleaning off all residual adhesive from molding.
N
No action
Compressor assembly
Fit a replacement spring and spring seat to the piston by twisting and pushing
and check that they are both secure.
Apply a 2-mm diameter sphere of
compressor lubricant
to the spring seat and
insert the piston; spring first, into the compressor body.
Carefully place the front cylinder housing over the piston, locating the boss in
the front cylinder housing recess and orientated to align the exhaust valve
above the outlet port.
Check that the front cover gasket is undamaged and supple. If it is not,
replace it. Fit the gasket over the pins on the underside of the cylinder head.
Carefully place the cylinder head and the gasket onto the compressor body
(with the serial number label uppermost), being careful to locate the pins in
the cylinder head in their corresponding holes in the compressor body.
Insert the cylinder head screws into the holes and tighten them up evenly using
a Phillips #2 screwdriver. DO NOT EXCEED 4.5lb/in (5.2kg/cm) TORQUE.
Apply the green dot to the air filter plate to identify the compressor has been
fitted with upgrade parts.
WHEN CONNECTING TO THE POWER SUPPLY, BEWARE OF
EXPOSED METAL PARTS AT HIGHER THAN LINE VOLTAGE.
Refit the compressor back into the controller.
Connect the controller to line voltage.
Test the compressor according to any method detailed in compressor
performance at the start of this sub-section.