rectangular hole. Before pressing down, make sure that the cut out is precisely lined
up with the recessed edges of the LCD window.
Smooth the membrane switch panel down, starting from the LCD window and
always working away from it, ensuring that there are no air bubbles.
Remove completely the backing paper on the new membrane fascia label (USA).
Align the fascia label accurately with the case recess and smooth the label down,
starting from the LCD window and always working away from it, ensuring that there
are no air bubbles.
Close the case.
Perform the complete controller functionality tests (see Section 2.2 or 2.3).
4.6 REPLACING THE COMPRESSOR
Disconnect the
compressor
from the
power supply assembly
at PL5 (see Figure 13),
and disconnect the compressor supply tube at the reservoir nipple.
Remove the large bore tube from the
tubing clips
on the side of the compressor
chassis, and lift it clear.
Remove the
moisture drain assembly
from the clip on the underside of the
compressor chassis.
Remove the screws, which hold the compressor chassis to the processor chassis, and
the screws that hold the compressor chassis to the controller case. Tilt and lift the
compressor assembly out of the case.
Remove the screws that hold the compressor to the chassis, and remove the tube
from the compressor outlet. Remove the compressor from the chassis, taking care
not to damage the
impact foam
.
Fit the compressor supply tube to the
compressor exchange
or new
compressor
.
Position the compressor within the chassis, pass the connecting leads and tube
through the grommets, and secure with the nuts, bolts, and washers.
BE CAREFUL
not to over tighten the nuts as this may damage the rubber mountings.
Put the compressor chassis assembly back into the controller case with the rubber
mountings pointing toward the fuses. Secure the chassis to the case with four screws
and washers, and secure to the processor chassis with two screws and washers.
Connect the tube to the reservoir and connect the lead to the power supply.
Replace the large bore tube and the moisture drain into the tubing clips.
Close the case.
Perform the complete controller functionality tests (see Section 2.2 or 2.3).