32
Perform the controller safety tests to confirm the ground resistance and leakage
current (see Sections 2.2.11 and 2.2.12).
Perform the complete controller functionality tests (see Section 2.2 or 2.3).
4.16 REPLACING THE POWER CORD
This is done without opening the case.
Remove the
power cord retainer
, and remove the
power cord
from the
power inlet
.
Insert the new power cord into the inlet socket, and secure with the power cord
retainer.
Perform the controller safety tests to confirm the ground resistance and leakage
current (see Sections 2.2.11 and 2.2.12).
4.17 REPLACING THE POWER INLET
Remove the
power cord retainer
and remove the
power cord
from the
power inlet
. Pull
off the three connectors from the rear of the inlet.
Use a screwdriver to collapse the retaining clips on each side of the inlet and push it
through from the inside of the case.
Figure 10: Arrangement of wires on the power inlet.
Power inlet viewed from inside of rear case
Line brown
Neutral Blue
Ground
Green/Yellow