Page 6 of 57
2.
SAFETY INSTRUCTIONS
The autoclave has unique characteristics. Please read and understand the
operation instructions before first operation of the autoclave. This manual
includes instructions guidance provided by the manufacturer: how to operate
the autoclave, the door safety mechanism, and the hazards involved in
circumventing safety means and how to select an adequate sterilization
program.
Make sure that you know where the main power switch is located.
Autoclave maintenance is crucial for the safe and effective function of the
device.
The daily B&D test, at the beginning of the working day, is part of the
preventive maintenance plan, along with the annual validation of the
sterilization processes that ensures appropriate sterilization conditions.
Never use the autoclave to sterilize liquids since it is not designed for that
purpose.
Below are the operating instructions – safety instructions:
1.
All autoclave users must receive training in proper usage from an
experienced employee. Every new employee must undergo a training
period under an experienced employee.
2.
Before use, check inside the autoclave chamber to ensure that no items
have been left from the previous cycle.
3.
Load
trays
in
such
a
way
as
to
allow
steam
to
move freely among all items.
4.
When sterilizing plastic materials, make sure that the item can withstand
sterilization temperature. Plastic that melts in the chamber is liable to
cause a great deal of damage.
5.
Individual
glass
bottles shall be placed on a tray.
6.
On closing the device door, make sure it is properly locked before
activating. Verify that
DOOR OPEN
symbol is replaced by the load
number.
7.
Verify once again that you have chosen the appropriate sterilization
program.
8.
Open the door slowly to allow steam to escape and wait 1 minute before
you remove the load.
9.
When removing the trays it is recommended to use the tray handle or
wear heat resistant gloves.
10.
Once a month, ensure that the safety valve is operating.
11.
A certified inspector must perform a periodic pressure chamber safety
test according to the local regulations.
12.
Once annually, or more frequently, effectiveness tests must be performed,
i.e., calibration and validation.
13.
Make
sure
there
are
no
leaks,
breaks,
blockages,
whistles
or
strange
noises.
14.
Perform maintenance operations as instructed. The owner of the
autoclave is responsible to perform the maintenance operations.
15.
Notify the person in charge immediately of any deviation from the
normal
functioning of the device.
16.
Protective equipment and clothes and other safety instructions should be
implemented in accordance with local and national regulations and/or
rules!