Page 23 of 57
Error
Number
Error
Name
Message Description
Corrective Action
Err:012 Manual
Stop
This error # is displayed and
cycle is aborted after the
STOP key is pressed.
Err:013 Air Valve
Error
This error # is displayed after
the air valve has been open for
3 minutes and the chamber is
unable to reach atmospheric
pressure.
Replace air filter.
Err:014 Low
Vacuum
The error # is
displayed if the
chamber cannot reach the
required vacuum conditions
after 20 minutes.
Verify that the door
is properly closed.
Perform a new
cycle.
If the problem
persists, call the
technician.
Err:015 Low Heat
This error # is displayed during
the Pre-Vacuum and Heating
stages if the system can't reach
the required pressure or
temperature conditions after 20
minutes.
Check that the
autoclave is not
overloaded.
If the problem
persists, call the
technician.
Err:016
Press
Sensor
Cutout
This error # and symbol are
displayed if the chamber
pressure
sensor
is
disconnected or out of range.
Call the technician.
Err:017
Pt2 out
This error # and symbol are
displayed when the Jacket
Temperature sensor Pt2 is
disconnected or out of range.
Call the technician.
Err:018
Press
Sensor
Cutout
This error # and symbol are
displayed if steam generator
pressure
sensor
is
disconnected
or out of range.
Call the technician.
Err:020 No Gen
water
This error # is displayed if the
generator electrode has no
water for a specific time.
This error # may appear the
first time the autoclave is
used and, in this case, does
not indicate any malfunction.
Make sure the
mineral-free
reservoir is full
Perform a new
cycle.
If the problem
persists, call the
technician.
Err:021 Test Fail
This error # is displayed if
one of the test programs fails.
Check for leakage
from the door.
Dry
the
chamber
and
try a second time.