Page 53 of 57
11.6.
Cleaning water strainer
CAUTION!
Before proceeding, make sure that there is no pressure in the
chamber.
CAUTION!
Do not touch the strainer shortly after operation.
Touching the hot strainer’s cover may cause severe injuries.
If maintenance operation is performed while strainer cover is hot,
use heat resistant gloves to avoid injuries.
1.
Turn the unit on momentarily to allow the door to be opened. Then
turn the unit off to stop the chamber from preheating.
2.
Remove the trays and the tray holder.
3.
Remove the strainer element mounted on the bottom-rear of the
chamber.
4.
Clean the strainer with water using a nylon or brass bristle brush if
necessary.
5.
Reinstall the strainer element.
6.
Reinstall the tray holder and the tray.
Strainer element
The strainer element is located
on the back of the chamber