41
9.2
Replacing the Air Filter (Models EA, EKA)
To facilitate drying the instruments with the door of the chamber closed,
models EA and EKA are equipped with an air compressor and HEPA
filter (0.2
µ
m). During the drying stage the compressor draws air through
the HEPA filter and forces the circulation of that air through the heated
chamber to remove moisture from the wrapped instruments. The filtration
of the air is performed by the bacteriological filter. Frequency of
replacement will be determined depending on the usage of the autoclave
and the surrounding environment.
The filter is mounted in an opening on the right sidewall of the autoclave
enclosure, this is to allow easy access for replacing it, (see picture below).
To replace the filter proceed as follows:
1.
Remove the filter cover (
2
) by turning the cover counterclockwise
until it is released.
2.
Remove the filter (5) from the filter cover by pulling the filter apart
from the filter cover
3.
Cut the tie wrap (
4
) fixing the flexible tube (3) connecting the filter
to the copper pipe, and pull off the filter.
4.
Replace the filter with a new one. Connect the filter (5) to the
flexible tube (3) and tighten it with a tie wrap (
4
).
5.
Connect the filter to the filter cover by pressing the filter (5) into the
hole in the cover (1).
6.
Insert the filter to its place inside the autoclave and reassemble the
filter cover by turning it a ¼ turn clockwise. Verify that the cover is
fastened well in its place.
2
4
3
5
2
1
Содержание 1730E
Страница 2: ......
Страница 16: ...14 REAR VIEW EA EKA only Cut Out Thermostat Reset button...
Страница 18: ...16 FRONT PANEL KEYBOARD NEW TYPE BLUE Note See section 3 2 for a description of the Keyboard buttons...
Страница 41: ...39 MAINTENANCE INSTRUCTIONS...