27
PRINTER OUTPUT
DESCRIPTION
Autoclave No: 01
Number of the autoclave with respect to other units in the
same location.
Load number: 0005
Load number. Useful to determine when to clean the
chamber, (upon reaching 255, this number is reset to 0).
Operator: ___________
To be filled in manually by operator.
O.K.
Cycle completed successfully.
D20
220°F
00P
The time, temperature and pressure during drying.
-----------------------
E20
251°F
02P
The time, temperature and pressure during exhaust.
-----------------------
S20
273°F
31P
The time, temperature and pressure during sterilization.
*
*
Prints sterilization data every 1 minute.
*
S13
273°F
31P
The time, temperature and pressure during sterilization.
S12
273°F
31P
The time, temperature and pressure during sterilization.
----------------------
H08
231°F
10P
The time, temperature and pressure during heating.
H04
137°F
00P
The time, temperature and pressure during heating.
*
*
Prints heat up data every 4 minute.
*
H00
72°F
00P
The time, temperature and pressure during heating.
----------------------
MN
TEMP
PRES
DRY: 30min
Drying time for selected program.
TIME: 08min
Sterilization time for selected program.
TEMP: 273°F
Sterilization temperature for selected program.
PROG: PKG
Selected program: Unwrapped instruments
TIME: 15:12:06
Time sterilization cycle begun.
DATE: 07:31:00
Date sterilization cycle begun.
Version:
T01EAWP
Number and version of the program
Legend
MN
Time elapsed in minutes
H
Heating stage
S
Sterilization stage
E
Exhaust stage
D
Drying stage
P
psi
Содержание 1730E
Страница 2: ......
Страница 16: ...14 REAR VIEW EA EKA only Cut Out Thermostat Reset button...
Страница 18: ...16 FRONT PANEL KEYBOARD NEW TYPE BLUE Note See section 3 2 for a description of the Keyboard buttons...
Страница 41: ...39 MAINTENANCE INSTRUCTIONS...