ID551PN
Weighing Controller
9
[2.11 ] Auto Zero Range
0
–
Disable Auto Zero
0.1~9.9 -
0.1d - 9.9d, Enable
When set to >0, the auto zero tracking is within
±
10d
[2.12 ] Auto Zero Speed
0
–
Disable Auto Zero
0.1~9.9 -
Enable Auto Zero, Speed is
(0.1~9.9)d/second
[2.13 ] Tare
0
– Disable Tare
1
–
Enable Tare
[2.14 ] Application Mode
0
– Weight mode
1
–
Dynamometry mode
[2.A ]
[2.A.1 ] :
View zero adjustment ADC code
[2.A.2 ] :
View span adjustment ADC code
[2.A.3 ] :
View span adjustment weight value
[2.A.4 ] :
Change or Enter zero adjustment ADC code
[2.A.5 ] :
Change or Enter span adjustment ADC code
[2.A.6 ] :
Change or Enter zero adjustment weight value
2.3.3 3#Channel Scale
[3.1 ] Increment Size
0
(0.001),
1
(0.002),
2
(0.005),
3
(0.01),
4
(0.02),
5
(0.05),
6
(0.1),
7
(0.2),
8
(0.5),
9
(1),
10
(2),
11
(5),
12
(10),
13
(20),
14
(50),
15
(100)
[3.2 ] Capacity
Sets the maximum weighing of the scales, with default value 60
[3.3 ] Calibration Mode
2-P
:
two point calibration mode ,Only support
[3.4 ]
(Zero Point)&
[3.6 ]
(End
Point
)
CALFr E
:
calibration without weighing added
[3.4 ] Zero Point Adjustment
See appendix 1.
[3.6 ] End-Point Calibration
See appendix 1.