ID551PN
Weighing Controller
19
MODBUS-RTU1 & MODBUS-TCP1 – 2 # Scale
Register
Description
Operation
40011/12
2# Scale
:
Floating point display weight
R
40013/14
2# Scale
:
Gross Weight
R
40015
2#Scale Status
Bit 0 – 0(Two point calibration mode) 1
(
CalFree
)
Bit1 - 0
(
Disable Tare
)
1
(
Enable Tare
)
Bit 7 –CalFree Error
Bit 8
– In
Center of Zero
Bit 9
–
Under Zero
Bit 10
–
Over Capacity
Bit 11
–
Net
Bit 12
–
Motion
Bit 13
–
Power Up Zero OK
Bit 14
– Setup Mode
Bit 15
–
Weight Data OK
R
40016
2#Scale Demand
:
Bit 0 – 0-Disable 1-Enable MODBUS remote calibration
Bit 1 - 0
(
Unlock Keypad
)
1(
Lock Keypad)
Bit 11
–
0->1
Trigger Pushbutton Zero
Bit 12
–
0->1
Trigger Pushbutton Tare
Bit 13
–
0->1
Trigger Pushbutton Clear
Bit 14
–
Bit 15
–
0->1
Trigger Scale Re-Init All
R/W
40017
2#Scale Calibration Demand: 0
1
Trigger Zero Point
Calibration
,
0
3
Trigger End Point Calibration
Notice : Only on 2 point calibration mode, support End Point
Calibration
R/W
40018
2#Scale
Capacity
,
only input integer weights
R/W
40019
2# Scale
Increment Size Index
:
0
(0.001),
1
(0.002),
2
(0.005),
3
(0.01),
4
(0.02),
5
(0.05),
6
(0.1),
7
(0.2),
8
(0.5),
9
(1),
10
(2),
11
(5),
12
(10),
13
(20),
14
(50),
15
(100)
R/W
40020
2# Scale :
Calibration Weight
,
only input integer weights
R/W