3.4 MODBUS-RTU1 & MODBUS-TCP1 Data Format

 

 

MODBUS-RTU1 & MODBUS-TCP1 – 1# Scale 

Register

 

Description

 

Operation

 

40001/2 

1# Scale

 

Floating point display weight

 

40003/4 

1# Scale

 

Gross Weight

 

40005 

1#Scale Status 

Bit 0 – 0(Two point calibration mode)     1

CalFree

 

Bit1 -    0

Disable Tare

     1

Enable Tare

 

 

Bit 7 – 1…….. CalFree Error   

Bit 8

 – In 

Center of Zero

             

Bit 9

 –

Under Zero

 

Bit 10

 –

Over Capacity

           

Bit 11

 –

Net

 

Bit 12

 –

Motion

              

Bit 13

 –

Power Up Zero OK

 

Bit 14 

– Setup Mode          

Bit 15

 –

Weight Data OK

 

 

 

 

 

40006 

 

 

1#Scale Demand

 

Bit 0 – 0-Disable  1-Enable MODBUS remote calibration 

Bit 1 -  0

Unlock Keypad

    1(

Lock Keypad)

 

Bit 11

 –

0->1

 

Trigger Pushbutton Zero

 

Bit 12

 –

0->1

 

Trigger Pushbutton Tare

 

Bit 13

 –

0->1

 

Trigger Pushbutton Clear

 

Bit 14

 – 

Bit 15

 –

0->1

 

Trigger Scale Re-Init All

 

 

 

R/W 

40007 

1#Scale  Calibration Demand:  0

1

  Trigger Zero Point  

Calibration

  0

Trigger End Point Calibration

 

Notice : Only on 2 point calibration mode, support End Point 

Calibration

 

R/W 

40008 

1#Scale 

Capacity

,

 only input integer weights 

 

R/W 

40009 

1#Scale 

Increment Size Index

(0.001), 

(0.002), 

(0.005),  

3

 (0.01), 

4

 (0.02), 

5

 (0.05), 

(0.1), 

(0.2), 

(0.5),   

9

 (1),   

10

 (2), 

11

 (5), 

12 

(10),     

13 

(20),    

14 

(50),     

15

 (100) 

R/W 

40010 

1# Scale : 

Calibration Weight

  only input integer weights

 

R/W 

 

 

 

 

 
 
 
 

 

Содержание ID551PN

Страница 1: ...1 ID551PN WEIGHING CONTROLLER OPERATION MANUAL 2017 03 Rev 1 00 ...

Страница 2: ... 3 6 Network Configuration 13 2 3 7 Maintenance 14 Chapter 3 Serial Port Application 15 3 1 Serial Port and Ethernet continuous output weight format 15 3 2 Continuous output A D Code format 16 3 3 Demand Input Output 16 3 4 MODBUS RTU1 MODBUS TCP1 Data Format 18 3 5 MODBUS RTU2 MODBUS TCP2 Data Format 22 3 6 PROFINET Data Format 24 Chapter 4 Hardware 26 4 1 Interface Specification 26 4 2 Housing S...

Страница 3: ...ID551PN Weighing Controller 3 Change Log Version Change Date V1 00 1 st Version 2017 03 ...

Страница 4: ... function Pushbutton Zero Auto Zero Pushbutton Tare Motion Checking Keyboard Lock UnLock MODBUS PROFINET remote calibration function Calibration 2 point Calibration or Cal Free Calibration Local Calibration or Remote Calibration via MODBUS RTU MOUBUS TCP PROFINET PLC communication Support PC software and Andriod phone APP for parameters upload download and save and calibrate scale weight display P...

Страница 5: ...ess enter setting menu Confirm input value Confirm exit Cursor indication Functionality PLC Indicates the communication status of ModbusRTU ModbusTCP Profinet flicker offline light ModbusRTU ModbusTCP Profinet At least one communication is normal 0 Light Scale in center of zero 1 4 division Light Scale in motion mode Net ON Scale in Net Mode 1 2 3 4 The current display and indicator represent chan...

Страница 6: ...port 1 4 Zero Point 1 6 End Point F1 1 Scale Configuration CALFrE calibration without weighing added 1 4 Zero Point Adjustment See appendix 1 1 6 End Point Calibration See appendix 1 1 7 CalFree Block 1 7 1 Load Cell Capacity Input the load cell capacity 1 7 2 Number of total load cells Input total number of load cells in this scale Setup Menu F2 2 Scale Configuration F3 3 Scale Configuration F4 4...

Страница 7: ...is 2 5 10 20 30 40 50 of capacity 1 10 Motion Check 1 A 1 A 1 View zero adjustment ADC code 1 A 2 View span adjustment ADC code 1 A 3 View span adjustment weight value 1 A 4 Change or Enter zero adjustment ADC code 1 A 5 Change or Enter span adjustment ADC code 1 A 6 Change or Enter zero adjustment weight value 3 0d 0 Disable motion check 0 1 9 9 Enable motion check 1 11 Auto Zero Range 0 Disable ...

Страница 8: ...k 2 7 1 Load Cell Capacity Input the load cell capacity 2 7 2 Number of total load cells Input total number of load cells in this scale 2 7 3 Number of working load cells Input the number of working load cells in this scale 2 7 4 Load Cell Sensitivity Input the sensitivity of load cell in general It is 2mV V or 3mV V 2 7 5 Sensitivity calibration A test weight is loaded to calibrate the sensor sen...

Страница 9: ... View span adjustment weight value 2 A 4 Change or Enter zero adjustment ADC code 2 A 5 Change or Enter span adjustment ADC code 2 A 6 Change or Enter zero adjustment weight value 2 3 3 3 Channel Scale 3 1 Increment Size 0 0 001 1 0 002 2 0 005 3 0 01 4 0 02 5 0 05 6 0 1 7 0 2 8 0 5 9 1 10 2 11 5 12 10 13 20 14 50 15 100 3 2 Capacity Sets the maximum weighing of the scales with default value 60 3 ...

Страница 10: ...erence End Point 3 8 Filter 0 lowest 1 2 3 4 5 18 19 Highest 3 9 Pushbutton Zero Range 0 Disable Pushbutton Zero 2 5 10 20 30 40 50 zero scale range is 2 5 10 20 30 40 50 of capacity 3 10 Motion Check 3 0d 0 Disable motion check 0 1 9 9 Enable motion check 3 11 Auto Zero Range 0 Disable Auto Zero 0 1 9 9 0 1d 9 9d Enable When set to 0 the auto zero tracking is within 10d 3 12 Auto Zero Speed 0 Dis...

Страница 11: ... 4 4 Zero Point Adjustment See appendix 1 4 6 End Point Calibration See appendix 1 4 7 CalFree Block 4 7 1 Load Cell Capacity Input the load cell capacity 4 7 2 Number of total load cells Input total number of load cells in this scale 4 7 3 Number of working load cells Input the number of working load cells in this scale 4 7 4 Load Cell Sensitivity Input the sensitivity of load cell in general It ...

Страница 12: ...A 2 View span adjustment ADC code 4 A 3 View span adjustment weight value 4 A 4 Change or Enter zero adjustment ADC code 4 A 5 Change or Enter span adjustment ADC code 4 A 6 Change or Enter zero adjustment weight value 2 3 5 Serial Port 6 1 Node Address The input range is 1 99 apply to MODBUS RTU MODBUS RTU support 1 99 6 2 Baud Rate 0 Baud rate is 1200 1 Baud rate is 2400 2 Baud rate is 4800 3 Ba...

Страница 13: ... below configured 9 1 1 192 9 1 2 168 9 1 3 1 9 1 4 10 F 9 2 Submask Default 255 255 255 255 as below configured 9 2 1 255 9 2 2 255 9 2 3 255 9 2 4 0 F 9 3 Gateway Default 192 168 0 1 as below configured 9 3 1 192 9 3 2 168 9 3 3 1 9 3 4 1 F 9 4 LAN 0 Disable 1 TCP 1025 port continuous output 2 TCP 1025 port continuous output support command C T Z F 9 5 MODBUS TCP 0 Disable 1 MODBUS TCP1 2 MODBUS...

Страница 14: ... key to switch 1 CLY or 0 CLY Display 1 CLY Indicates that the station name exists then restart Display 0 CLY Indicates that the station name does not exists no need to clear F10 F F10 F 1 Password Default 0000 no password If not 0000 you need to enter a password F10 F 2 Keyboard Lock 0 Keyboard not locked 1 Lock Keyboard Under the keyboard lock state only the Switch key is useful Clear Tare and Z...

Страница 15: ...XXXX Notice A B C D E Byte 1byte 1byte 1byte 6bytes 6bytes Notice A SWA Status A B SWB Status B C SWC Status C D DisplayWeight 6 digits without decimal E Tare Weight 6 digits without decimal SWA Status A Bit 0 Bit2 Bit1 Bit0 Decimal Point Location 0 0 1 XXXXX0 0 1 0 XXXXXX 0 1 1 XXXXX X 1 0 0 XXXX XX 1 0 1 XXX XXX 1 1 0 XX XXXX 1 2 3 Bit4 Bit3 Build Code 0 1 X1 1 0 X2 1 1 X5 4 5 Always 1 6 Always ...

Страница 16: ...EN ODD Paritybit 3 2 Continuous output A D Code format Output Data Start 1 A D Code Separator 2 A D Code Separator 3 A D Code Separator 4 A D Code END Data 0x02 7 bytes comma 7 bytes comma 7 bytes comma 7 bytes 0x0D Notice A D Code is 7 bits ASCII Code less than 7 bits fill 0 the Separatoris comma 0x2C 3 3 Demand Input Output Both Serial Port Ethernet can be configured demand input output below co...

Страница 17: ...ID551PN Weighing Controller 17 NC Clear Tare NZ Zero Scale Description The above N 1 2 3 4 correspond to 1 2 3 4 Scale For example 1T is sent 1 Scale perform Tare operations ...

Страница 18: ...06 1 Scale Demand Bit 0 0 Disable 1 Enable MODBUS remote calibration Bit 1 0 Unlock Keypad 1 Lock Keypad Bit 11 0 1 Trigger Pushbutton Zero Bit 12 0 1 Trigger Pushbutton Tare Bit 13 0 1 Trigger Pushbutton Clear Bit 14 Bit 15 0 1 Trigger Scale Re Init All R W 40007 1 Scale Calibration Demand 0 1 Trigger Zero Point Calibration 0 3 Trigger End Point Calibration Notice Only on 2 point calibration mode...

Страница 19: ...t 0 0 Disable 1 Enable MODBUS remote calibration Bit 1 0 Unlock Keypad 1 Lock Keypad Bit 11 0 1 Trigger Pushbutton Zero Bit 12 0 1 Trigger Pushbutton Tare Bit 13 0 1 Trigger Pushbutton Clear Bit 14 Bit 15 0 1 Trigger Scale Re Init All R W 40017 2 Scale Calibration Demand 0 1 Trigger Zero Point Calibration 0 3 Trigger End Point Calibration Notice Only on 2 point calibration mode support End Point C...

Страница 20: ...t 0 0 Disable 1 Enable MODBUS remote calibration Bit 1 0 Unlock Keypad 1 Lock Keypad Bit 11 0 1 Trigger Pushbutton Zero Bit 12 0 1 Trigger Pushbutton Tare Bit 13 0 1 Trigger Pushbutton Clear Bit 14 Bit 15 0 1 Trigger Scale Re Init All R W 40027 3 Scale Calibration Demand 0 1 Trigger Zero Point Calibration 0 3 Trigger End Point Calibration Notice Only on 2 point calibration mode support End Point C...

Страница 21: ...t 0 0 Disable 1 Enable MODBUS remote calibration Bit 1 0 Unlock Keypad 1 Lock Keypad Bit 11 0 1 Trigger Pushbutton Zero Bit 12 0 1 Trigger Pushbutton Tare Bit 13 0 1 Trigger Pushbutton Clear Bit 14 Bit 15 0 1 Trigger Scale Re Init All R W 40037 4 Scale Calibration Demand 0 1 Trigger Zero Point Calibration 0 3 Trigger End Point Calibration Notice Only on 2 point calibration mode support End Point C...

Страница 22: ... 1 Lock Keypad Bit 11 0 1 Trigger Pushbutton Zero Bit 12 0 1 Trigger Pushbutton Tare Bit 13 0 1 Trigger Pushbutton Clear Bit 14 Bit 15 0 1 Trigger Scale Re Init All R W 40007 8 2 Scale Floating point display weight R 40009 10 2 Scale Gross Weight R 40011 2 Scale Status Bit 0 0 Two point calibration mode 1 CalFree Bit1 0 Disable Tare 1 Enable Tare Bit 7 CalFree Error Bit 8 In Center of Zero Bit 9 U...

Страница 23: ...1 0 1 Trigger Pushbutton Zero Bit 12 0 1 Trigger Pushbutton Tare Bit 13 0 1 Trigger Pushbutton Clear Bit 14 Bit 15 0 1 Trigger Scale Re Init All R W 40019 20 4 Scale Floating point display weight R 40021 22 4 Scale Gross Weight R 40023 4 Scale Status Bit 0 0 Two point calibration mode 1 CalFree Bit1 0 Disable Tare 1 Enable Tare Bit 7 CalFree Error Bit 8 In Center of Zero Bit 9 Under Zero Bit 10 Ov...

Страница 24: ...erved W5 Reserved W6 W11 is 2 Scale W12 W17 is 3 scale W18 W23 is 4 scale Instruction Request PLC ID551 Floating Point 1 Scale W0_1 W0_8 S0 S7 Read or Write Variable Index When W0_0 W0_8 changes and W0_0 is 1 Write will trigger a write command If W0_0 is read operation Read the corresponding variables in real time W0_0 0 Read 1 Write W0_9 W0_11 Reserved W0_12 1 lock keypad 0 unlock keypad W0_13 W0...

Страница 25: ... S0 S7 Read or Write Variable Index W0_9 The result of Write operation 0 success 1 Fail Reason W_1 W_2 1 Unable Write 2 Write error 3 Reserved W0_10 Motion Output 0 Invalid 1 valid W0_11 Net status 0 Gross 1 Net W0_12 Data Bit 1 W0_13 1 Under Zero W0_14 1 Over Capacity W0_15 1 Data OK W1 W2 write variable s value to ID551 DWORD Format W3 W4 Display Weight in DWORD Format W5_0 W5_7 Reserved W5_8 1 ...

Страница 26: ...n board SW 1 ON W M Sealed ON Calibration is Disabled OFF W M Sealed OFF Calibration is Enabled SW 2 Reserved keep OFF SW 3 ON Load Factory Default Values OFF Normal keep OFF for normal use SW 4 Reserved keep OFF Profinet RS232 485 AC power IN Keyboard Display 6 line Load Ce ll Conne ct 4 line Load Ce ll Conne ct ...

Страница 27: ...ID551PN Weighing Controller 27 4 2 Housing Size DIN Rail Housing ...

Страница 28: ...l lock Disable Calibration 255 Calibration Fail 0 Calibration Done Range point calibration 2 P X is 1 2 3 4 corresponding to the 4 channel 10 9 2 1 Calibration 100 Calibration Motion 200 Seal lock Disable Calibration 255 Calibration Fail 0 Calibration Done FX 6 CAL 2 N Y 9 8 1 0 SAVE End Point Calibrate Completely F1 5 CAL 0 N Y 9 8 1 0 SAVE Zero Adj Completely Go to F1 6 ...

Страница 29: ...ta error EE 100 Tare failed Disable Tare EE 101 Tare failed for scale in motion EE 102 Tare failed out of tare range EE 105 Zero failed for Zero is disabled EE 106 Zero failed for scale in motion EE 107 Zero failed out of zero range EE 108 No keys allowed for keypad is locked EE 109 PLC option board start failed EE 111 ADC Init failed please re power or contact factory EE no The product key does n...

Отзывы: