ID551PN
Weighing Controller
6
2.2 Top Menu
2.3 Setup
2.3.1 1#Channel Scale
[1.1 ] Increment Size
0
(0.001),
1
(0.002),
2
(0.005),
3
(0.01),
4
(0.02),
5
(0.05),
6
(0.1),
7
(0.2),
8
(0.5),
9
(1),
10
(2),
11
(5),
12
(10),
13
(20),
14
(50),
15
(100)
[1.2 ] Capacity
Sets the maximum weighing of the scales, with default value 60
[1.3 ] Calibration Mode
2-P
:
two point calibration mode ,Only support
[1.4 ]
(Zero Point)&
[1.6 ]
(End
Point
[F1] 1#Scale Configuration
)
CALFr E
:
calibration without weighing added
[1.4 ] Zero Point Adjustment
See appendix 1.
[1.6 ] End-Point Calibration
See appendix 1.
[1.7 ] CalFree Block
[1.7.1]
Load Cell Capacity
Input the load cell capacity
[1.7.2]
Number of total load cells
Input total number of load cells in this scale
[Setup]
Menu
[F2] 2#Scale Configuration
[F3] 3#Scale Configuration
[F4] 4#Scale Configuration
[F10]
Maintenance
Press Print Key for 2 seconds
[F9]
Network
[- - - -]
If have a
password,
please enter
password
[F6]Serial Port