3
7
DIAGNOSTICS MENU
(continued)
MENU ITEM
SOFTWARE
NAME ITEM
DESCRIPTION
EXHAUST
ANALOG OUTPUT
EXHAUST
AOUT
The
EXHAUST AOUT
item is used to verify the analog outputs are working. When this item is entered, the
number shown on the display indicates the last analog output value. The value displayed ranges from 0 to 255. The
value 255 corresponds to 0 volts (4 mA) output and 0 corresponds to 10 volts (20 mA) output. Pressing the
S
/
T
keys change the count on the display. Pressing the
S
key should increase the analog output. Pressing the
T
key
will decrease the analog output.
The
EXHAUST AOUT
function used in conjunction with a volt meter will verify the individual analog output is
operating correctly.
RESET THE
CONTROLLER TO
FACTORY
DEFAULT
SETTINGS
RESET TO
DEF
When this menu item is entered, the 8682 will prompt the user to verify that they want to do this by indicating NO.
Use the
S
keys change the display to YES, then press the
SELECT
key to reset the controller to its factory
defaults. Pressing the
MENU
key before the
SELECT
key will exit out of the menu item.
WARNING:
If YES is selected, the Model 8682 will reset all menu items to their factory default settings: The
controller will have to be reprogrammed and recalibrated after this operation is completed.
END
OF
MENU
The
END OF MENU
item informs you that the end of a menu has been reached. You can either scroll back up the
menu to make changes, or press the
SELECT
or
MENU
key to exit out of the menu.