Placement and Calibration of Sensors
Calibrating CO2 AND O2
EVM Series User Manual
61
5.
In the
Calibration
screen, press
▲
/
▼
Arrows
to select CO
2
or O
2
.
Selecting a sensor
Press Up or Down Arrow
CAL soft key
Figure 4-23: Selecting a calibration sensor
6.
Press the
CAL
soft key to open the CO
2
“Calibration/Calibrate” screen. (Allow at least 2.5 minutes
for CO
2
to stabilize.)
For the Set Points, -
zero
and
–span
soft keys will appear. Press either
Zero Cal
or
Span Cal
soft key. (See Table 4-2 for more information.)
Soft key Menu
Zero, Span or Exit
Figure 4-24: Calibration CO
2
ready screen
NOTE:
If “
–
” (a minus sign) appears in front of Zero or Span, this has not been calibrated.
Alternatively, the “
+
” (plus sign) signifies the cal point has been calibrated.
7.
Turn the regulator on. The regulator should have a flow rate of 1.0 L/min.
8.
Allow the level to stabilize. For CO
2
, wait at least two minutes to stabilize.
For Zero Cal, the screen will display Zero with a Set soft key.
For span cal, press
▲
/
▼
Arrows
on the keypad to adjust the appropriate level.
Содержание EVM SERIES
Страница 1: ...ENVIRONMENTAL MONITOR EVM SERIES USER MANUAL...
Страница 2: ......
Страница 6: ......
Страница 20: ...EVM Series Models Explained EVM Series User Manual 10 This page intentionally left blank...
Страница 52: ...Setting up Parameters via the Keypad Auto run EVM Series User Manual 42 This page intentionally left blank...
Страница 100: ...File System Menu Estimated Run Time EVM Series User Manual 90 This page intentionally left blank...
Страница 118: ...EVM Reports and Printing EVM Firmware Update EVM Series User Manual 108 This page intentionally left blank...
Страница 144: ...Warranty EVM Series User Manual 134 This page intentionally left blank...