Placement and Calibration of Sensors
Pump Flow Rate Calibration
EVM Series User Manual
53
To reset, press the down arrow repeatedly to select the gravimetric fields and press the
Reset
soft key. (If it is not displayed, press the
Altf
key.)
5.
Turn the pump on by pressing the
Pump
key.
6.
Optional step:
a.
It is optional to press the
Run/Stop
key to log the gravimetric filter sampling.
NOTE
: If in “run mode”, a clock appears and displays the overall run time.
b.
To view the particulate level, press the right arrow once from the start screen and the
particulate measurement screen will appear (noted as “PM”).
7.
After the gravimetric collection, turn the pump off and reference the
Gravimetric Mass
field
to view
the calculated particulates.
8.
Lastly, calculate the ratio between the value in the
Gravimetric Mass
field and the weight of the
collected sample.
9.
Enter in the correction factor via the
Particulate Setup
screen (see “
” section for
more details).
Pump Flow Rate Calibration
An accurate flow meter, such as a primary standard flowmeter calibrator, and inert tubing is required to
perform this calibration. The inert tubing is inserted on the dust air flow outlet (located on the back of the
housing).
NOTE
: You may want to send the instrument in for a yearly calibration, if you do not own a
flowmeter calibrator for this calibration procedure.
Air Flow Outlet
Used to connect a primary standard calibrator with
tubing
Figure 4-14: Pump flow rate calibration
Flow rate calibration
1.
Connect the flow meter (i.e, primary standard) and inert tubing to the EVM.
2.
Turn the EVM on by pressing
On/Off/Esc
.
3.
Select the
Calibration
menu by pressing the
Down Arrow
until it is selected and press
Enter
.
4.
Select
Flow Rt
(flow rate) by repeatedly pressing
▲
/
▼
Arrows
key until it appears.
Содержание EVM SERIES
Страница 1: ...ENVIRONMENTAL MONITOR EVM SERIES USER MANUAL...
Страница 2: ......
Страница 6: ......
Страница 20: ...EVM Series Models Explained EVM Series User Manual 10 This page intentionally left blank...
Страница 52: ...Setting up Parameters via the Keypad Auto run EVM Series User Manual 42 This page intentionally left blank...
Страница 100: ...File System Menu Estimated Run Time EVM Series User Manual 90 This page intentionally left blank...
Страница 118: ...EVM Reports and Printing EVM Firmware Update EVM Series User Manual 108 This page intentionally left blank...
Страница 144: ...Warranty EVM Series User Manual 134 This page intentionally left blank...