61
Chapter 6
Service and Maintenance
Recalibration Interval
TSI recommends that the PortaCount
®
Pro Respirator Fit Tester be
cleaned and recalibrated annually. Try to arrange your fit testing
schedule to allow for annual factory recalibration. If you have a fit
testing schedule that is seasonal or you expect heavy usage during
an upcoming period (such as an outage), TSI recommends that you
have the PortaCount Pro fit tester serviced prior to that time. TSI's
Customer Service Department provides fast service in order to
minimize your down time. Contact TSI on the Internet at
or send e-mail to
Status Messages
Two status messages can be displayed by the PortaCount Pro
Respirator Fit Tester.
Low Particle
Message
The PortaCount Pro fit tester is factory
programmed not to allow fit testing when the
ambient particle concentration is below 1000
particles/cm
3
for the Model 8030 or 8038 when
testing masks with 99% efficiencies or greater,
and 30 particles/cm
3
for the Model 8038 when
testing masks with <99% efficiencies. If the
ambient sample in Fit Test Mode is below these
levels
, the “Low Particle” message appears on
the touch screen display and the fit test is
automatically terminated.
This message appears in Fit Test Mode only, for
one of the reasons described below. Consult the
Содержание PORTACOUNT PRO+ 8038
Страница 2: ......
Страница 10: ...4 Introduction and Overview This page intentionally left blank ...
Страница 22: ...16 Chapter 2 This page intentionally left blank ...
Страница 30: ...24 Chapter 3 This page intentionally left blank ...
Страница 82: ...76 Chapter 7 This page intentionally left blank ...
Страница 86: ...80 Appendix A This page intentionally left blank ...
Страница 90: ...84 Appendix B This page intentionally left blank ...
Страница 94: ...88 Appendix D This page intentionally left blank ...
Страница 96: ...90 Appendix E This page intentionally left blank ...
Страница 100: ...94 Appendix F This page intentionally left blank ...
Страница 101: ...95 Appendix G The Rainbow Passage ...
Страница 102: ...96 Appendix G This page intentionally left blank ...
Страница 106: ......