Conducting a Fit Test with the PortaCount Pro Respirator Fit Tester
27
4. Set the Storage Cap and Alcohol Cartridge down on a clean
surface to prevent contamination.
5. Open a bottle of alcohol. Invert the bottle and insert the nozzle
end into the Alcohol Fill Capsule as far as possible to make
certain alcohol is not inadvertently sprayed anywhere except into
the Capsule.
6. Squeeze alcohol into the Alcohol Fill Capsule until it is even with
or slightly above the fill-line.
7. Recap the alcohol bottle immediately.