Chapter 4
8. Make certain the alcohol cartridge is clean. If it has been
contaminated, refer to the
chapter and
follow the instructions to replace it. If it is clean, insert the Alcohol
Cartridge into the Alcohol Fill Capsule and turn the capsule
clockwise until it locks in place.
9. Set the Alcohol Fill Capsule down and wait at least 2 minutes
while the alcohol wick inside the Alcohol Cartridge soaks up
alcohol.
10. Remove the Alcohol Cartridge from the Capsule and gently shake
it to allow excess alcohol to drip back into the Alcohol Fill
Capsule. Stop when excess alcohol is no longer dripping; it is not
necessary to wait until the outside surface of the Alcohol
Cartridge is dry.
11. Insert the Alcohol Cartridge into the Cartridge Cavity of the
PortaCount Pro Respirator Fit Tester. It should slide in with little
effort.
DO NOT FORCE IT
(see Note below). As you approach full
insertion, firmly twist the Alcohol Cartridge clockwise until it locks
into position.
Содержание PORTACOUNT PRO+ 8038
Страница 2: ......
Страница 10: ...4 Introduction and Overview This page intentionally left blank ...
Страница 22: ...16 Chapter 2 This page intentionally left blank ...
Страница 30: ...24 Chapter 3 This page intentionally left blank ...
Страница 82: ...76 Chapter 7 This page intentionally left blank ...
Страница 86: ...80 Appendix A This page intentionally left blank ...
Страница 90: ...84 Appendix B This page intentionally left blank ...
Страница 94: ...88 Appendix D This page intentionally left blank ...
Страница 96: ...90 Appendix E This page intentionally left blank ...
Страница 100: ...94 Appendix F This page intentionally left blank ...
Страница 101: ...95 Appendix G The Rainbow Passage ...
Страница 102: ...96 Appendix G This page intentionally left blank ...
Страница 106: ......