About Respirators and Fit Testing
21
For respirators with an outer mesh
, install the probe normally
as if the mesh was not present. The sampling probe and push nut
will seal properly right through the mesh.
2. Slide the sampling probe onto the piercing tool.
As the pointed
end of the piercing tool is very sharp, be extremely careful
when handling it!
3. Choose a location on the mask that is in front of the person's
nose/mouth region. Avoid seams and folds in the mask.
4. Pierce the mask at the selected location, using the piercing tool
with the loaded sampling probe.
Be sure to pierce the mask
from the inside!
Sampling Probe
Содержание PORTACOUNT PRO+ 8038
Страница 2: ......
Страница 10: ...4 Introduction and Overview This page intentionally left blank ...
Страница 22: ...16 Chapter 2 This page intentionally left blank ...
Страница 30: ...24 Chapter 3 This page intentionally left blank ...
Страница 82: ...76 Chapter 7 This page intentionally left blank ...
Страница 86: ...80 Appendix A This page intentionally left blank ...
Страница 90: ...84 Appendix B This page intentionally left blank ...
Страница 94: ...88 Appendix D This page intentionally left blank ...
Страница 96: ...90 Appendix E This page intentionally left blank ...
Страница 100: ...94 Appendix F This page intentionally left blank ...
Страница 101: ...95 Appendix G The Rainbow Passage ...
Страница 102: ...96 Appendix G This page intentionally left blank ...
Страница 106: ......