![Troxler RoadReader Plus 3450 Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/troxler/roadreader-plus-3450/roadreader-plus-3450_manual-of-operation-and-instruction_1172544073.webp)
5–
Gently place the control unit inside the gauge and tighten the four
screws in the corners of the keypad. Turn the gauge on. If the gauge
displays
Baseboard Not Connected
, then the ribbon cable is not
connected properly. Check between the baseboard and the control
unit connector again.
If the gauge displays
Incorrect Baseboard ID
, then the control
unit is connected to the wrong gauge. The calibration constants for a
gauge are stored in its control unit; therefore, the control unit must
be returned to the gauge from which it was removed. To ensure that
this occurs, the control unit checks an electronic ID on the base. If
this error is displayed, return the control unit to the correct gauge.
Содержание RoadReader Plus 3450
Страница 7: ...Model 3450 vii NOTES...
Страница 13: ...xii NOTES...
Страница 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Страница 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 53: ...3 28 NOTES...
Страница 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 181: ...Appendix G 4 NOTES...
Страница 187: ...Appendix H 6 NOTES...
Страница 199: ...Index 12 NOTES...
Страница 200: ...Model 3450 Warranty 1 NOTES WARRANTY...