STAT TEST
Erratic readings, or readings that seem to fluctuate, may indicate a
problem with the gauge. If the readings are suspect, perform a
stat
test
, or statistical stability test, to validate the normal operation of
the gauge.
A stat test consists of 20 one-minute counts. From the 20 counts, the
gauge calculates the standard deviation. This standard deviation is
compared to a theoretical standard deviation value. Ideally, this ratio
should be 0.354, with acceptable limits from 0.225 to 0.465. The
gauge is considered unstable if the ratio is outside these limits and
the stat test will fail. If the stat test fails, ensure that the gauge setup
and testing was conducted correctly, and take an additional test. If
the second test fails, call the nearest Troxler service center for
assistance.
To access the
Stat Test
menu, press the
〈
SPECIAL
〉
key. At the
Special
functions menu, access the
Gauge Status/Test
menu by
pressing
〈
2
〉
. At the
Gauge Status/Test
menu, press
〈
2
〉
. The gauge
displays the
Stat Test
menu:
The following text provides details on the stat test function.
1- Take STAT Test
2- Review STAT Test
3- Print STAT Data
Press ESC To Exit
C. T
RO
UB
LE
SH
OO
TIN
G
Содержание RoadReader Plus 3450
Страница 7: ...Model 3450 vii NOTES...
Страница 13: ...xii NOTES...
Страница 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Страница 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 53: ...3 28 NOTES...
Страница 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 181: ...Appendix G 4 NOTES...
Страница 187: ...Appendix H 6 NOTES...
Страница 199: ...Index 12 NOTES...
Страница 200: ...Model 3450 Warranty 1 NOTES WARRANTY...