Model 3450
Appendix C–17
SERIAL DEVICE NOT CONNECTED!
This error occurs when the serial input signals from printing device
are incorrect.
Solution
: Check for correct serial cable. Printer must be on line.
(Serial printer only.)
SYSTEM ERROR! SYSTEM DATA MAY BE LOST
This error occurs when the software system variables have been
corrupted.
Solution
: The software attempts to correct the problem, but gauge
setup may be lost. If problems continue to occur, call the nearest
Troxler Service Center.
WARNING! LEAK TEST IS DUE. LAST LEAK TEST DATE:
XX/XX/XXXX
This error occurs when the leak test interval (default: 6 months) has
elapsed since the last reset of the leak test date.
Solution
: Perform a leak test and reset the leak test date to current
date.
WARNING! SOURCE ROD NOT AT BS
This error occurs when the software expects the source rod to be at a
particular position and detects that it is not.
Solution
: Check source rod position to ensure it is at the proper
position. Recalibrate depth strip. Call the nearest Troxler Service
Center.
C. T
RO
UB
LE
SH
OO
TIN
G
Содержание RoadReader Plus 3450
Страница 7: ...Model 3450 vii NOTES...
Страница 13: ...xii NOTES...
Страница 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Страница 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 53: ...3 28 NOTES...
Страница 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 181: ...Appendix G 4 NOTES...
Страница 187: ...Appendix H 6 NOTES...
Страница 199: ...Index 12 NOTES...
Страница 200: ...Model 3450 Warranty 1 NOTES WARRANTY...