Model 3450
6–21
The gauge displays:
Enter the number of measurements the gauge is to average. Press the
〈
ENTER
〉
key.
With the source rod in the backscatter position, begin the
measurement by pressing the
〈
START
〉
key. The gauge displays the
progress.
After taking the reading, the gauge displays the results. To continue,
press the
〈
ENTER
〉
key. Follow the gauge prompts to complete the
calibration.
Edit a Partial Thin Layer Special Calibration
If a partial calibration was saved earlier, the gauge will display a
P
beside the option entered before. Select the value to be completed.
If the value already entered is selected, the gauge prompts the
operator to overwrite. To overwrite the partial calibration, press the
〈
YES
〉
key and go to page 6–18. To use the stored partial
calibration, press the
〈
NO
〉
key and select the other value.
To take measurements, see page 6–18. To enter the true density, see
page 6–20.
Input Number of Thin
Lift Averages
_ _ (3-20)
Then Press ENTER
Ready For
Count 1 Of xx
Place in BS Then
Press START
6
. O
FF
SE
TS
&
C
AL
IB
RA
TIO
NS
Содержание RoadReader Plus 3450
Страница 7: ...Model 3450 vii NOTES...
Страница 13: ...xii NOTES...
Страница 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Страница 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 53: ...3 28 NOTES...
Страница 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 181: ...Appendix G 4 NOTES...
Страница 187: ...Appendix H 6 NOTES...
Страница 199: ...Index 12 NOTES...
Страница 200: ...Model 3450 Warranty 1 NOTES WARRANTY...