Model 3450
4–
✓
After the count time, the gauge displays the measurement
results.
where:
%MA
= percent Marshall
WD
= wet density
%VOIDS = 100
×
1 – WD
VOIDLESS
(displayed only when enabled)
If a project is enabled (see the
section on
page 5–5), press the
〈
STORE
〉
key to store the measurement(s).
To scroll through the results, use the arrow keys. Return to the
Ready
screen by pressing the
〈
ESC
〉
key, or press the
〈
START
〉
key to make another measurement.
✓
Lift the gauge from the test site using the source rod handle.
This returns the source rod to the
SAFE
position. When not
taking readings, always keep the source rod in the
SAFE
position.
%MA = xxxxx %
WD = xxxxx pcf
%VOIDS = xxxxx
4. U
SIN
G T
HE
G
AU
GE
Содержание RoadReader Plus 3450
Страница 7: ...Model 3450 vii NOTES...
Страница 13: ...xii NOTES...
Страница 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Страница 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Страница 53: ...3 28 NOTES...
Страница 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Страница 181: ...Appendix G 4 NOTES...
Страница 187: ...Appendix H 6 NOTES...
Страница 199: ...Index 12 NOTES...
Страница 200: ...Model 3450 Warranty 1 NOTES WARRANTY...