![Trolex Sentro 8 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/trolex/sentro-8/sentro-8_user-manual_1171050088.webp)
88 TX9165-UM-EN-07
7.3 Service Replacement Sensing
Modules
Service replacement Sentro eModules and
rModules can be supplied by our Product
Support Department on a regular basis.
Contact:
for assistance.
Simply insert the replacement eModule or
rModule into the Sentro 8 and return the
original for calibration using the envelope
provided.
7.4 Replacing a Sentro Sensing
Module
Identify the sensing module to be replaced.
To prevent a
Fault
condition being generated
when the sensing module is removed it is
necessary to
Exclude
the sensing module.
From the
Base Screens
, use the navigation
keypad, scroll to the required sensing
module, press and hold down the left arrow
on the navigation keypad.
Use the navigation keypad, scroll to
Status
and select
OK
.
Select
OK
to access the
Exclude Menu
.
Use the navigation keypad, scroll to
Yes
and
select
OK
.
Use the navigation keypad, scroll to
Set
or
Quit
and select
OK
.
www.trolex.com