70 TX9165-UM-EN-07
Checkpoint
The standard configuration for a Sentro
8 has rModules fitted in locations E to H.
It is possible that eModules are fitted in
locations E to H but this is a factory fitted
configuration. If you are unsure which
modules you have fitted to your Sentro 8,
look at the
Sensor Inputs Overview
for
channels group E to H and see if gas types
are displayed. If gas types are displayed
you have eModules fitted in channel group
E to H. If you are still unsure contact:
for assistance.
4.5.1 Scaling
The various characteristics of the analogue
input signal scale values can be configured.
To Access Scaling
From the
Sentro rModule Setup Menu
, use
the navigation keypad, scroll to
Scaling
and
select
OK
.
If a
Security Code
has been enabled you will
be prompted to enter it, enter the
Security
Code
and select
OK
.
Sig Fig
The position of the decimal point can be
shifted along the displayed number to the
best
Significant Figure
to suit the particular
sensor signal being monitored and its
optimum measuring range.
This can be used to eliminate digit flicker and
redundant decimal places.
www.trolex.com
Содержание Sentro 8
Страница 1: ......
Страница 2: ......
Страница 8: ...8 TX9165 UM EN 07 1 4 Dimensions 1 4 1 TX9165 Sentro 8 Sensorstation Four Gas Ports www trolex com...
Страница 37: ...TX9165 User Manual TX9165 UM EN 07 37 4 Setup and Calibration 4 1 Controls and Indicators www trolex com...
Страница 38: ...38 TX9165 UM EN 07 www trolex com...
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......