TX9165 User Manual
TX9165-UM-EN-07 61
Check
The screen displays the
Span
value
of calibration gas that the Sentro 8
Sensorstation expects to be used during
calibration.
Check the
Span
value of gas displayed
against the value on your calibration gas
cylinder.
If the two values match then proceed to the
Calibrate Span
procedure.
If they do not match then
Change
the value
as follows:
Use the navigation keypad, scroll to
Change
and select
OK
.
With
Expected
highlighted select
OK
.
Use the navigation keypad, scroll up or down
on the first digit to the required
Span
value
on your calibration gas cylinder. Scroll right
to the second digit. Repeat the above for the
second and third digit in
Span
and select
OK
.
Scroll to
Accept
and select
OK
.
Continue with
Calibrate Span
as described
in the following section.