4.4.1 Sentro eModules Calibrate
Checkpoint
The standard configuration for a Sentro 8
has eModules fitted in locations A, B, C and
D. It is possible that eModules are fitted in
locations E, F, G and H but this is a factory
fitted configuration. If you are unsure which
modules you have fitted to your Sentro 8,
look at the
Sensor Inputs Overview
for
channels group E to H and see if gas types
are displayed. If gas types are displayed
you have eModules fitted in channel group
E to H. If you are still unsure contact:
for assistance.
Checkpoint
Ensure the area where the equipment is
being calibrated is well ventilated. Observe
appropriate Health and Safety legislation
and applicable local procedures when
handling test gases.
Checkpoint
When calibrate is selected the eModule
will output a fixed value of 0.00. This
prevents the possibility of false alarm
signals being activated in the monitoring
equipment during the calibration process
(20% will be output for an oxygen sensor).
Fit a gas hood to the corresponding port of
the eModule being calibrated. Connect the
gas application tube from the clean air test
gas cylinder to the gas hood fitted to the
eModule being calibrated.
www.trolex.com
Содержание Sentro 8
Страница 1: ......
Страница 2: ......
Страница 8: ...8 TX9165 UM EN 07 1 4 Dimensions 1 4 1 TX9165 Sentro 8 Sensorstation Four Gas Ports www trolex com...
Страница 37: ...TX9165 User Manual TX9165 UM EN 07 37 4 Setup and Calibration 4 1 Controls and Indicators www trolex com...
Страница 38: ...38 TX9165 UM EN 07 www trolex com...
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......