Appendix B
Appendix B – Medical Gas Alarm Specifications
Operating Ambient Temperature range
: +10C(50F) to +50C(122F)
Storage Temperature:
-20C(-4F) to +85C(185F)
AC Input:
120 - 240 volts AC -
50-60 Hz
Input Fuse:
5 amp input AC line fuse protects the input wiring to power supply
Power Consumption:
45W maximum
Pressure Measurement Accuracy:
0-30 inHg transducer +/-1%
Vacuum, Gas Evacuation
0-100 PSIG tran/-1%
Oxygen, Nitrous Oxide, Medical Air, Carbon Dioxide
0-250PSIG tran/-1%
Nitrogen
Dimensions
Rough-in Box
-
All dimensions are in inches and cover the basic box only
(mounting flange excluded)
Logic one module
8.125W x 11.875H x 4.000D
Logic two modules
13.625W x 11.875H x 4.000D
Logic four modules
19.625W x 11.875H x 4.000D
Front Panel
Logic one module
10.945W x 13.125H x 1.250D
Logic two modules
14.870W x 13.125H x 1.250D
Logic four modules
22.720W x 13.125H x 1.250D
Transducers:
Housing dimensions: 1.990W x 1.990H x 3.625 Length
including inlet fittings