11.3 Conductivity Electrode Care and Maintenance
● Ensure that the conductivity electrode is thoroughly washed with distilled water after each
use.
● If there is a build-up of solids inside the measurement area of the cell, these should be
removed very carefully with a cotton bud soaked in solvent, taking care not to touch the
metal parts of the inner cell.
11.4 DO Probe Care and Maintenance
● Always keep probe membrane moist.
● If dissolved oxygen probe is not in use for long periods, unscrew membrane cap and rinse
the probe's cathode, anode and the membrane with deionised water, then soak up residual
water with filter paper, and re-install the probe.
Cathode and Anode
Membrane cap