4.5 Temperature Compensation
For accurate results, either manual or automatic temperature compensation must be enabled
before measurement or calibration.
4.6 Automatic Temperature Compensation:
● Insert the 6-pin connector on the temperature probe into the corresponding connector on
the instrument, ensuring that the lugs are aligned. Once pushed firmly into place, tighten the
waterproof locking collar.
● The “ATC” indicator will show on the display; the meter is now switched to automatic
temperature compensation mode.
4.7 Manual Temperature Compensation:
4.7.1. DO NOT connect the temperature probe to meter.
4.7.2. Press and hold the MODE key for 3 seconds to enter temperature setting mode.
4.7.3. Press
▲ or ▼ key to set the temperature of the sample.
4.7.4. Press ENTER key to confirm, the meter returns to measurement mode. Setting is
completed.
In the temperature setting mode, pressing and releasing the
▲ or ▼ key will increase or
decrease the temperature value by 0.1. Pressing and holding the
▲ or ▼ key will increase or
decrease the temperature value by 1.