Trace2o Ltd
Rev1.0
Notes
For best results, rinse vials thoroughly between with deionised water. Ensure that the
outside of the vials are clean, dry and free from fingerprints. Always handle vials by the lid
where possible. For deep cleaning of vials, immerse in Sodium Hypochlorite solution for 1
hour and rinse with deionised water.
The tablets should be added direct from the foil avoiding contact with hands or surfaces.
The sample should have a pH value of between 6.2 and 6.5 after addition of the reagent.
High Calcium ion content and/or high conductivity can lead to turbidity of the sample and
therefore incorrect measurements – a solution is available if necessary.
Interferences can occur from oxidising agents such as Bromine and Ozone.