5.2 LCD backlight 

By pressing this button 

 you will activate the LCD backlight. With each operation the light will switch on for 

about 7 seconds. When switching off the gauge the light will also automatically switch off. 

5.3 Automatic power shut down 

If the instrument is not operated for over 2 minutes it will automatically switch off. 

6 CALIBRATIONS 

We recommend annual calibration. For calibration, send the instrument, together with a RMA form* to TQC, 
Molenbaan 19, 2908 LL  Capelle aan den IJssel, NL.  

*You can download the RMA form here: 

http://www.tqc.eu/en/service/repairs-calibrations/

6

 

REPLACING BATTERIES 

When the icon

 flashes, please replace the batteries before taking any new measurements. Turn the 

instrument off by pressing the ON/OFF button. Open the battery lid and replace them with 3 pcs of AAA alkaline 
batteries. 

7

 

MAINTENANCE 

š

 

Though robust in design, this instrument is precision-machined. Never drop it or knock it over  

š

 

Always clean the instrument after use.  

š

 

Clean the instrument using a soft dry cloth. Never clean the instrument by any mechanical means such as a 
wire brush or abrasive paper. This may cause, just like the use of aggressive cleaning agents, permanent 
damage.  

š

 

Do not use compressed air to clean the instrument.  

š

 

Always keep the instrument in its case when not in use. 

š

 

We recommend annual calibration 

8

 

DISCLAIMER 

The right of technical modifications is reserved. 

Backlight on

Backlight off 

Содержание LTD7015

Страница 1: ...will give accurate readings to an accuracy of 1 This unit is not suitable for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring freq...

Страница 2: ...ng temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored The transducer is made of propyle...

Страница 3: ...eted After calibration the sound velocity will remain the same as previous selected A calibration should be made the every time the probe or batteries are replaced This is very important to make sure...

Страница 4: ...When the icon flashes please replace the batteries before taking any new measurements Turn the instrument off by pressing the ON OFF button Open the battery lid and replace them with 3 pcs of AAA alk...

Страница 5: ...or otherwise is correct we have no control over either the quality or condition of the product or the many factors affecting the use and application of the product Therefore unless we specifically agr...

Страница 6: ...e for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring frequency 5MHz 2 5MHz Measuring range Accuracy Sound velocity Display Resolut...

Страница 7: ...utor for a replacement probe Working temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored...

Страница 8: ...small amount of coupling gel on the sample block and then place the probe on the sample block When the display shows 4 0mm the calibration is completed After calibration the sound velocity will remain...

Страница 9: ...ss MODE again to store After selecting your memory unit you can start measuring again While measuring the memory unit will flash and the new measurement will delete the stored value When your measurem...

Страница 10: ...the axis of the pipe to be measured For small pipes two measurements should be performed one with the axis of the pipe perpendicular the other one parallel The smallest of the 2 values displayed shou...

Страница 11: ...calibrating the gauge with a reference block However this table is just for reference sometimes the sound velocity will be different caused by different physical chemical factors The sound velocity of...

Страница 12: ...ve recess surface incorrect calibration sample blocks or the inner flaw of materials 8 Coupling gel Coupling gel causes the high frequency ultrasonic waves transmitting between the probe and material...

Страница 13: ...t in use We recommend annual calibration 9 DISCLAIMER The right of technical modifications is reserved The information given in this manual is not intended to be exhaustive and any person using the pr...

Отзывы: