5. Multilayer-composite material 
It is impossible to read out the thickness of uncoupled multilayer’s as the ultrasonic wave cannot go through 
into the uncoupled space. Furthermore, the sonic wave cannot go through in to the composite material at an 
even speed. So ultrasonic reflect principles cannot be applied for measuring multilayer’s and composite 
material. 
 
6. Influences from oxidized materials 
For some metals such as aluminium, a layer of oxide can be generated on the surface. Although the oxidised 
layer combines with the substrate tightly, the sonic wave does need to travel through 2 different materials so it 
will lead to incorrect readings. With a thicker oxidised layer, more deviation will occur. Please calibrate the 
instrument with a reference block of the same material as the sample. Obtain the thickness of the sample block 
by using a micrometer/calliper. 

 

7.

  

Any

 

abnormal readings 

An operator should be capable  to distinguish any abnormal readings as a result of erosive, recess surface, 
incorrect calibration sample blocks or the inner flaw of materials. 
 
8.

 

Coupling gel

 

Coupling gel causes the high frequency ultrasonic waves transmitting between the probe and material. 
Choosing incorrect coupling or wrong operation may cause errors or pour coupling which results in failure to 
measure. The coupling gel should be used in a proper way, normally, a small drop of gel is sufficient. It is 
important to use proper coupling gel, low viscosity gel (as supplied) is suitable for smooth surfaces. For rough, 
vertical, aluminium surfaces, high viscosity liquids like glycerine and lubrication grease is applicable. These 
kinds of agents are normally available in the local market but you can buy them also from a local distributor.

 

 
Sound Velocity table: 
 

Material Velocity 

m/s Material Velocity 

m/s 

Aluminium  6320 Acetate 

resin 2670 

Zinc 4170 

Phosphor 

3530 

Silver 3600 

Turpentine 

4430 

Gold 3240 Glass 5440 

Tin 3230 

Incoloy 

alloy 

5720 

Iron/steel 5900 

Magnesium 

6310 

Brass 

4640 Monel 

alloy 6020 

Copper 4700  Nickel  5630 

SUS 

5790 

Steel 4330 (mid) 

5850 

Acrylic resin 

2730 

Steel 330 

5660 

Water 

(20°C) 

1480 Titanium 6070 

Glycerin  1920 Zirconium 4650 

Soluble glass 

2350 

Nylon 

2620 

 
 

Содержание LTD7015

Страница 1: ...will give accurate readings to an accuracy of 1 This unit is not suitable for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring freq...

Страница 2: ...ng temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored The transducer is made of propyle...

Страница 3: ...eted After calibration the sound velocity will remain the same as previous selected A calibration should be made the every time the probe or batteries are replaced This is very important to make sure...

Страница 4: ...When the icon flashes please replace the batteries before taking any new measurements Turn the instrument off by pressing the ON OFF button Open the battery lid and replace them with 3 pcs of AAA alk...

Страница 5: ...or otherwise is correct we have no control over either the quality or condition of the product or the many factors affecting the use and application of the product Therefore unless we specifically agr...

Страница 6: ...e for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring frequency 5MHz 2 5MHz Measuring range Accuracy Sound velocity Display Resolut...

Страница 7: ...utor for a replacement probe Working temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored...

Страница 8: ...small amount of coupling gel on the sample block and then place the probe on the sample block When the display shows 4 0mm the calibration is completed After calibration the sound velocity will remain...

Страница 9: ...ss MODE again to store After selecting your memory unit you can start measuring again While measuring the memory unit will flash and the new measurement will delete the stored value When your measurem...

Страница 10: ...the axis of the pipe to be measured For small pipes two measurements should be performed one with the axis of the pipe perpendicular the other one parallel The smallest of the 2 values displayed shou...

Страница 11: ...calibrating the gauge with a reference block However this table is just for reference sometimes the sound velocity will be different caused by different physical chemical factors The sound velocity of...

Страница 12: ...ve recess surface incorrect calibration sample blocks or the inner flaw of materials 8 Coupling gel Coupling gel causes the high frequency ultrasonic waves transmitting between the probe and material...

Страница 13: ...t in use We recommend annual calibration 9 DISCLAIMER The right of technical modifications is reserved The information given in this manual is not intended to be exhaustive and any person using the pr...

Отзывы: