WHAT’S IN THE BOX?
Ultrasonic wall thickness gauge
Coupling gel 60ml
Probe 10mm - 2.5MHz
Probe 10mm - 5.0MHz
Calibration plate
3
OPERATION INSTRUCTIONS
Operation area condition: This gauge can measure all samples which are equal to or bigger than the probes
surface area.
Curved surface condition: When materials have a curved surfaces like a boiler wall or tubing, the curvature
radius must be equal to or more than 10mm and the wall thickness must be equal to or more than 3mm.
Roughness condition: This gauge is applicable for a wide range of materials. In most cases the supplied
probe makes very accurate readings. However if the roughness is too big due to rust etc. an error may
occur. In this case please try to minimise the roughness or select the 2.5MHz probe. If the probe is worn-
out, please contact your local distributor for a replacement probe.
Working temperature condition: Material and sound velocity will change along with temperature. In
normal measurement conditions environment temperature impact can be ignored. The probe is made of
propylene material. Considering the protection of the probe and its precision we recommend a surface
temperature of the work piece should not be over 60
o
C otherwise the probe could be damaged.
Operation temperature: 0 - 40
o
C
Relative humidity: <90% RH
Hardware / Work piece/ Material temperature: < 60
o
C
Do not apply on heavily vibrating / erosive material
Avoid impact and moisture
4
PREPARATIONS
1. Turning on your instrument
Connect the probe with the main unit, press ON/OFF to turn on the instrument. The display shows a full screen
for 0.5 sec with backlight, then a vertical bar is displayed twice circularly, meanwhile the main display area
shows 0000 to 9999 progressively. After that the LCD displays the last applied sound velocity with registered
memory.
Full screen
Last applied sound velocity & unit no.