3 |

2

.

Sound velocity selection and adjustment

.

Press “VEL” key to enter the sound velocity interface, press again, you can now select the velocity data by 
pressing         or         the  word “VEL” will be blinking on the display. If you need to adjust the sound velocity, 
press the “VEL” key once more, the “VEL” and “m/s” will keep blinking. Then increase or decrease the sound 
velocity to the number you want by pressing        or        again. The data will be automatically stored into the 
current sound velocity memory unit. After this, press to confirm new sound velocity and the icons will stop 
blinking. 

3

.

Calibration.

If necessary this step should be repeated; Press the “ZERO” key for 2 seconds, after a short time “CAL” will 
appear on the display, this indicates the unit is in calibration mode. Put a small amount of coupling gel on the 
sample block and then place the probe on the sample block. When the display shows 4.0mm the calibration is 
completed. After calibration the sound velocity will remain the same as previous selected. A calibration should 
be made the each time the probe or batteries are replaced. This is very important to make sure measurements 
will be accurate. 

5

 

PERFORM A MEASUREMENT 

5.1 Thickness measurement 

Put some coupling gel on the material surface, press the ultrasonic probe on the material surface to measure. 
Please make sure there is good coupling, which will be indicated with this symbol         appearing on the display.  
If this symbol does not appear or flashes it means there is not a good coupling. The reading on the display is the 
measured value. After removing the probe, the reading will be held.  

Calibration status 

Calibration accomplished

Good coupling measurement 

Measurement accomplished 

velocity adjustment 

velocity revision 

Содержание LTD7015

Страница 1: ...will give accurate readings to an accuracy of 1 This unit is not suitable for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring freq...

Страница 2: ...ng temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored The transducer is made of propyle...

Страница 3: ...eted After calibration the sound velocity will remain the same as previous selected A calibration should be made the every time the probe or batteries are replaced This is very important to make sure...

Страница 4: ...When the icon flashes please replace the batteries before taking any new measurements Turn the instrument off by pressing the ON OFF button Open the battery lid and replace them with 3 pcs of AAA alk...

Страница 5: ...or otherwise is correct we have no control over either the quality or condition of the product or the many factors affecting the use and application of the product Therefore unless we specifically agr...

Страница 6: ...e for cast iron due to its big crystalloid composition 1 1 Specifications Measuring method Ultrasonic pulse echo Measuring frequency 5MHz 2 5MHz Measuring range Accuracy Sound velocity Display Resolut...

Страница 7: ...utor for a replacement probe Working temperature condition Material and sound velocity will change along with temperature In normal measurement conditions environment temperature impact can be ignored...

Страница 8: ...small amount of coupling gel on the sample block and then place the probe on the sample block When the display shows 4 0mm the calibration is completed After calibration the sound velocity will remain...

Страница 9: ...ss MODE again to store After selecting your memory unit you can start measuring again While measuring the memory unit will flash and the new measurement will delete the stored value When your measurem...

Страница 10: ...the axis of the pipe to be measured For small pipes two measurements should be performed one with the axis of the pipe perpendicular the other one parallel The smallest of the 2 values displayed shou...

Страница 11: ...calibrating the gauge with a reference block However this table is just for reference sometimes the sound velocity will be different caused by different physical chemical factors The sound velocity of...

Страница 12: ...ve recess surface incorrect calibration sample blocks or the inner flaw of materials 8 Coupling gel Coupling gel causes the high frequency ultrasonic waves transmitting between the probe and material...

Страница 13: ...t in use We recommend annual calibration 9 DISCLAIMER The right of technical modifications is reserved The information given in this manual is not intended to be exhaustive and any person using the pr...