23
ANNEX A | INSTALLATION OF THE VACUUM BED
After unpacking the TQC automatic Film applicator the rubber seals for the vacuum bed need to
be placed. Without these seals the vacuum can NOT be created.
Remove all tape holding the Perforated- or Double Channel Vacuum Bed. Take great care to
remove all the tape.
Remove the vacuum bed from the base bed, and stow it in a safe place.
Take the two black rubber rings from the package and differentiate in size*.
Place the smallest of the rings (A4)* around the inner supports and the largest (A3) around the
outer supports, as shown in below image (detail A).
Gently lower down the vacuum bed (B or C). Take care not to disturb the placement of
the rubbers.
Turn the device on and follow the instruction to start a new test, using an A3 test chart, however
hold at the step where the vacuum pump turn on.
Place an A3 test chart and press it down. It should be sucked to the vacuum bed. If not grab the
sides of both the base plate and the vacuum bed and press them together.
When the pitch of the vacuum pump changes and sounds softer the vacuum seal is complete.
Continue the run and afterwards you are ready.
*
When installing double channel only place A3 ring
4A
C
B
3A
Detail
A
scale 1:1
EN
Содержание AB3075
Страница 25: ...25 ANNEX C INSTALLATION OF THE LAMP Detail A scale 1 1 EN...
Страница 39: ...39 TQC GRINDOMETER TOOL FOR AFA AB3075 ANNEX F TQC GRINDOMETER TOOL FOR AFA EN...
Страница 41: ...41 ANNEX F GRINDOMETER TOOL FOR AFA Weight WWW TQC EU AB3075 WWW TQC EU AB3075 100 g 805 g 114 g EN...
Страница 42: ...42 ANNEX F GRINDOMETER TOOL FOR AFA Tool Operation 1 WWW TQC EU AB3075 WWW TQC EU AB3075 EN...
Страница 44: ...44 ANNEX F GRINDOMETER TOOL FOR AFA Height adjustment 3 WWW TQC EU A B 5 mm WWW TQC EU AB3075 EN...
Страница 46: ...46 ANNEX F GRINDOMETER TOOL FOR AFA After drawdown 5 WWW TQC EU AB3075 EN...
Страница 47: ...47 TQC HARDNESS PEN HOLDER TOOL FOR AFA AB3090 ANNEX G TQC HARDNESS PEN HOLDER FOR AFA EN...
Страница 48: ...48 ANNEX G TQC HARDNESS PEN HOLDER FOR AFA Dimensions 60 mm 140 mm 25 mm 52 mm 23 mm EN...
Страница 49: ...49 ANNEX G TQC HARDNESS PEN HOLDER FOR AFA Weight 500 g 2220 g EN...
Страница 50: ...50 EN ANNEX G TQC HARDNESS PEN HOLDER FOR AFA Tool Mounting...
Страница 52: ...52 EN ANNEX G TQC HARDNESS PEN HOLDER FOR AFA TEST PANEL PLACEMENT Vacuum models X Vacuum hole Mask Position...
Страница 54: ...54 EN...
Страница 55: ...55 EN...