Page 8
General Specifications
Temperature Compensation : Automatic, 0 to 100
o
C,
or manual for pH, 0 to 99.0
o
C.
Cond. Sensor Span Range : k=0.1 Sensor : k=0.075 to k=0.133
k=1.0 Sensor : k=0.75 to k=1.33
k=10 Sensor : k=7.5 to k=13.3
pH Asymmetry Range
: -1.00 to 1.00 pH
pH Slope Range
: 85.0 to 105%
Temp. Sensor Offset Range : -10.0
o
C to +10.0
o
C
Auto Standard Recognition : Cond.
: 150
µ
S/cm, 1413
µ
S/cm, 2.76 mS/cm,
12.88 mS/cm, 58.0 mS/cm
TDS
: 69.5 ppM, 2.00 ppK, 8.00 ppK, 36.0 ppK
pH
: 4.00, 6.88, 7.00, 9.23
Memory
: 150 readings including date and time
Automatic Logging
: User-set for one reading every 1 to 90 seconds or
minutes.
RS232 Output (optional)
: 300, 1200 & 9600 baud.
8 bits, no parity, 1 stop bit, XON/XOFF Protocol.
Clock
: Calendar clock displays date, month, hours, minutes
& seconds.
Year is recorded in memory and transmitted to
optional RS232 port, but is not displayed.
Battery Saver
: On : Auto switch-off after 5 minutes
Off : Continuous use
Bar Graph display of battery charge level.
Readout of battery voltage available for
troubleshooting.
Good Laboratory Practices : Date, Time and Value of last Conductivity, TDS,
pH and Temperature calibration are stored, and can
be recalled or sent to the optional RS232 port at any
time.
Power
: 6V NiCad Rechargeable Battery for approx
50 hours operation.
Dimensions
: 187 x 110 x 51 mm
Mass
: Instrument only
: Approx 440g
Full Kit
: Approx 1.7kg
Environment
Temperature
: 0 to 45
o
C
Humidity
: 0 to 90 % R.H.