Page 27
11.6 Data Format
A.
Data is returned to the RS232 port by the
WP-81T
in the following
format when requested by a PC with the ?D or ?R commands
(section 11.5):
LLLL
ss
DDDDDDUUU
ss
PPPPPPuuu
ss
TTTTTToCm
ss
dd/mm/yy
ss
hh:mm:ss
<cr>
or B.
Data is sent to the RS232 port by the
WP-81T
in the following format
when it is sent by the
WP-81T
using the Print function (section 9.4) or
the Instant Send function (section 11.2):
LLLL
ss
DDDDDDUUU
ss
PPPPPPuuu
ss
TTTTTToCm
ss
dd/mm/yy
ss
hh:mm:ss
<cr><lf>
where: LLLL
is the Log Number. Maximum 4 characters, right justified.
The
WP-81T
sends a Zero for instant readings (section
11.2)
ss
is one space.
DDDDDD
is the Conductivity or TDS Data. Maximum 6
characters, right justified.
UUU
is the unit description, either “uS
ss
”
, “mS
ss
”
, “ppM” or
“ppK” (where
ss
is 1 space).
ss
is one space.
PPPPPP
is the pH Data. Maximum 6 characters, right justified.
uuu
is the unit description, sent as “pH
ss
” (where
ss
is 1 space).
ss
is one space.
TTTTTT
is the Temperature Data. Maximum 6 characters, right
justified.
oCm
is the Temperature unit description. The
WP-81T
sends
“oC
ss
” for real temperature data (where
ss
is one space),
or “oCm” when manual temperature compensation is being
used.
ss
is one space.
dd/mm/yy
is the date, month and year data.
ss
is one space.
hh:mm:ss
is the hours, minutes and seconds data.