Page 20
8.2 Failed Calibration
If calibration has failed, the GLP function will reset the date and time to zero. The
WP-81T
still shows the results of the last successful calibration.
eg:
Cond Zero 0.00uS
k=1.00
@
00/00/00 00:00
@
00/00/00 00:00
Asy 0.10pH
Slope 100.0%
Temp Offset 1.0
o
c
@
00/00/00 00:00
@
00/00/00 00:00
@
00/00/00 00:00
Note that these calibration values are still used if further measurements are taken
without recalibrating.
8.3 Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a printer or
PC via the RS232 port. This function is available only when the optional RS232
port is fitted.
1.
Switch the meter on.
2.
Ensure that the
WP-81T
RS232 cable is connected to the instrument and to
the printer or PC.
3.
Send the GLP information to the RS232 port:
→
F4:Setup
→
F3:GLP
→
F3:Print
4.
The GLP information is sent to the RS232 port in formatted ASCII text.
eg:
WP81T V1.0 R1234 @ 31/12/97 12:00
Conductivity Zero= 0.00uS @ 31/12/97 11:00
Conductivity k= 1.00 @ 2.76mS @ 31/12/97 11:10
TDS Zero= 0.00ppM @ 31/12/97 11:20
TDS k= 1.00 @ 36.0ppK @ 31/12/97 11:30
pH Asy= 0.00pH @ 31/12/97 11:40
pH Slope= 100.0% @ 31/12/97 11:50
Temperature Offset= 1.0oC @ 31/12/97 12:00
ENDS