Page 28
C.
GLP information is returned as 8 lines terminated by a carriage return.
When using the “?G” command (section 11.5), the computer must
respond with a character after receiving each line.
eg:
WP81T V1.0 R1234 @ 31/12/97 12:00
Conductivity Zero= 0.00uS @ 31/12/97 11:05
Conductivity k= 1.00 @ 2.76mS @ 31/12/97 11:10
TDS Zero= 0.00ppM @ 31/12/97 11:15
TDS k= 1.00 @ 36.0ppK @ 31/12/97 11:20
pH Asy= 0.00pH @ 31/12/97 11:25
pH Slope= 100.0% @ 31/12/97 11:30
Temperature Offset= 1.0oC @ 21/12/97 11:35
ENDS
12. Battery Saver Function
The
WP-81T
is equipped with a battery saver function. If no button has been
pressed for five minutes, the unit beeps and flashes the display for 20 seconds, and
then shuts off. This function can be switched off for continuous use.
To enable or disable the battery saver function:
1.
Switch the meter on.
2.
Select Battery Saver Set-up
→
F4:Setup
→
F1:Batt
3.
The battery saver menu is now displayed.
eg:
Batt Saver
F1:OFF
100%
→
→
F2:ON
The arrow indicates the current selection.
The bar graph and percentage indicate the approximate level of charge in the
battery.
4.
Press
to disable the battery saver function for continuous use.
Press
to enable the battery saver function. The meter will switch itself off
if no key has been pressed for five minutes.
Press
to quit the battery saver menu and retain the current setting.
5.
NOTE:
For troubleshooting purposes, the battery volts can also be displayed
in the battery saver menu. Press
to display battery volts.
The
symbol flashes when the battery volts drops below 5.60 volts.
At 5.00 volts the meter turns itself off.