Page 18
7.4 Manual Temperature Setting
If the Conductivity/TDS/Temperature sensor is not connected, and a temperature
sensor is not used in its place, the temperature of the sample solution must be set
manually for accurate pH measurements. A separate thermometer will be required
for this.
NOTE:
The Conductivity sensor has a separate sensor built in for automatic
temperature compensation for Conductivity and TDS.
1.
Switch the meter on.
2.
Measure the temperature of the sample.
3.
Select Temperature Calibration
→
F1:Cal.
→
F3:Temp
4.
The current temperature setting is now displayed.
eg:
→
→
25.0
←
←
Man Temp
↑↑↓↓
:Set F1:Save
5.
Press the
and
keys until the display shows the temperature of the
sample.
6.
Press the
key to save the temperature value.
Alternatively, press the
key to quit and retain the current setting.