pH and mV Troubleshooting, continued...
Inaccurate readings,
even when
calibration is
successful.
Reference junction blocked.
Clean reference junction, as per
instructions supplied with the
electrode.
Displays 7.00 for
all solutions.
Electrical short in connector.
1.
Check connector. Replace if
necessary.
2.
Replace electrode.
Displays 4-5 pH for
all solutions.
Glass bulb or internal stem
cracked.
Replace electrode.
18.4 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Displays “
OVR
o
C
”
when electrode is
plugged in.
1.
Faulty electrode.
2.
Faulty instrument.
Fit new electrode, part number
122201.
Return to factory for repair.
Temperature
inaccurate and
cannot be
calibrated.
1.
Faulty connector.
2.
Faulty electrode.
3.
Faulty instrument.
Check the connector and
replace if necessary.
Fit new electrode, part number
122201.
Return to factory for repair.