Page 22
9. Notepad Function
9.1 Recording Readings into the Notepad
To record readings into the Notepad memory:
1.
Press
in normal display mode. The display should now look like this:
eg:
2.76mS
7.00pH
or
2.00ppK
7.00pH
F1: 1
12:00:00
F1: 1
12:00:00
2.
If you now press
, the Conductivity/TDS, pH, Temperature, Date and
Time will be recorded into the notepad, and labeled as reading number 1.
3.
Repeat steps 1 & 2 as often as required. The maximum number of readings
that can be stored in the Notepad is 150.
9.2 Recalling Records from the Notepad
To recall records from the Notepad onto the
WP-81T
display:
1.
Select the Notepad menu
→
F2:Notepad
2.
Select Recall from the menu
F1:Recall
3.
Record number 1 is now displayed,
eg:
2.76mS
7.00pH
25.0
o
c
1 F2:Clk
4.
Press
to alternatively display the date and time or the data for this record.
Press
to move forward through the records.
Press
to move backward through the records.
Press and hold the
or
keys to roll rapidly through the readings.
9.3 Erasing Records from the Notepad
To erase all records from the Notepad:
1.
Select the Notepad menu
→
F2:Notepad
2.
Select Erase from the menu
F2:Erase
3.
The
WP-81T
now asks if you are sure that you wish to erase all records.
Erase, You Sure?
F1:Yes F2:No
4.
Press
to erase all records from the Notepad
Press
to quit without erasing the records from the Notepad.