17.4 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Temperature inaccurate and
cannot be calibrated.
1. Faulty connector.
2. Faulty Dissolved Oxygen or
Temperature sensor,
whichever is being
calibrated.
3. Faulty Dissolved Oxygen
cable if Dissolved Oxygen
sensor is being used for
Temperature readout.
Check the connector and replace if
necessary.
Return Dissolved Oxygen or Temperature
sensor for repair, or replace sensor.
Return Dissolved Oxygen cable for repair
or replace cable.
Meters displays “
Probe
”
instead of Temperature
data.
Meter has been Temperature-
calibrated on ED1 and then a
YSI sensor is connected, or vice
versa.
Re-calibrate the Temperature reading on the
new sensor. Meter only keeps Temperature
calibration data for one Dissolved Oxygen
sensor.
Displays flashing “
M
” when
Temperature and/or
Dissolved Oxygen sensor
plugged in.
1. Faulty instrument socket.
2. Faulty Dissolved Oxygen or
Temperature sensor,
whichever is being
calibrated.
3. Faulty Dissolved Oxygen
cable if Dissolved Oxygen
sensor is being used for
Temperature readout.
Return the instrument to the TPS factory for
service.
Return Dissolved Oxygen or Temperature
sensor for repair, or replace sensor.
Return Dissolved Oxygen cable for repair
or replace cable.
Reading does not change in
any sample being measured
Auto Stability Function has
frozen display (notice the ).
Turn Auto Stability Function OFF or re-
start sampling (see section 7).