6.2 Temperature Calibration Notes
1. Temperature calibration information is retained in memory when the
smart
CHEM
-D
is
switched off, even when the power supply is removed. This information can be recalled later
using the GLP function (see section 9).
2. Temperature does not need to be re-calibrated unless the Temperature or Dissolved Oxygen
sensor is replaced or the meter is initialised.
3. The
smart
CHEM
-D
keeps the Temperature calibration data for the Temperature sensor and
one type of Dissolved Oxygen sensor separately so these can be swapped as required.
6.3 Calibration Messages
1. If a temperature calibration has been successfully performed, the
smart
CHEM
-D
will display
the following message and the offset of the sensor.
Calibration OK
Offset=0.1
O
C
2. If a temperature calibration has failed, the
smart
CHEM
-D
will display the following message
and the failed offset value of the sensor.
Calibration Failed
Offset=11.0
O
C
3. The
smart
CHEM
-D
has an allowable Offset range of -10.0 to +10.0
O
C
. If calibration fails due
to the Offset being outside these limits, then please consult the Troubleshooting guide (section
17.4) for possible remedies.
4. If neither a Temperature sensor nor a Dissolved Oxygen sensor is connected to the
smart
CHEM
-D
, and a temperature calibration is attempted, then the following message will be
displayed.
Must plug in either
Temp or Oxygen probe